Olympus A50002A electronic laparoscope image abnormality

Release time:

2025-03-12 09:36

Machine Type:Olympus Electronic Laparoscope

Machine Model:A50002A

1. Malfunction Phenomenon

Image blur abnormality

a3f2215b6cd0a534af3d5a824f078361.png
2. Fault Detection
Check for damage and destruction. Cable breakage, aging of the light guide, damage to the light guide, damage to the CCD tip glass.
3. Fault Determination and Analysis
Long-term use leads to cable breakage, aging and damage of the light guide and light guide; the tissue protein remaining on the surface of the insertion tube is dehydrated and adhered to the outer skin during disinfection, and accumulates over time to form crystals, causing damage to the CCD component.
4. Fault Handling
Replace the light guide; repair the cable; replace the CCD; replace the light guide.
5. Repair and On-Machine Effect
35547cc8d799c5495ad340d0e1ae2172.png

6. Endoscope Tips

After daily use

(1) After use, the endoscope should be strictly cleaned, disinfected, dried, and leak-tested according to the instructions to extend the service life of the endoscope.

(2) After use, it should be immediately cleaned with cleaning solution, and the channel should be cleaned with a special cleaning brush before being placed in the soaking solution for soaking. At any time, the soft part of the endoscope should not be held or cleaned together with hard and sharp accessories. When cleaning with cloth, use clean and soft non-woven fabric to prevent foreign objects from contaminating the endoscope body.

(3) After cleaning the endoscope, all pipes should be blown dry, and the lens surface should be wiped clean in the same direction with lens paper to keep the lens clean and bright.

(4) Regularly check the endoscope for leaks to prevent water ingress from causing imaging failures.

 

 

 

 

 

 

 

This information comes from the Internet. Please contact us to delete it if there is any infringement.