Two Cases of Philips HD15 Ultrasound Machine Repair

Release time:

2025-03-12 09:38

Malfunction 1

1.1 Malfunction Phenomenon

The PHILIPS HD15 machine experienced a problem with the small display screen not showing any image, and the control panel's indicator lights occasionally failed to illuminate.

1.2 Malfunction Analysis and Troubleshooting

Upon arriving on-site and observing the machine's malfunction, it was noted that the fault had minimal impact on clinical diagnostic applications, thus the machine remained in use. All indicator lights on the control panel were illuminated, and the buttons functioned normally. Inquiry with the department staff about the frequency of the malfunction revealed that the machine occasionally exhibited a failure where all indicator lights would not illuminate after startup, but the buttons remained functional. This occurred infrequently. The main malfunction was the small display screen not showing any image. The HD15 machine has a simple operating display panel above the control panel, with nine small display screens. The displayed content corresponds to eight parameter adjustment knobs and one page-turning control button on the control panel. This simple operating display panel malfunction is a common fault with the HD15 machine. After replacing the panel, the display returned to normal. After replacing the control panel, it was used for one month without any recurrence of the indicator lights failing to illuminate.

Malfunction 2

2.1 Malfunction Phenomenon

After powering on the machine, the monitor displayed “NO SIGNAL POWER SAVING MODE DIGITAL”.

2.2 Malfunction Analysis and Troubleshooting

On-site observation of the malfunction after powering on the machine showed that both the monitor and workstation displayed “NO SIGNAL POWER SAVING MODE DIGITAL”, thus ruling out a monitor malfunction. Measurements of the power supply voltages during machine startup were normal, and the same malfunction occurred after replacing the power supply, ruling out a power supply problem. Analysis of the machine's startup operation indicated that the fault originated within the PC case. After removing the PC case cover and restarting the machine, the motherboard fan spun normally and there were normal self-test sounds. The control panel indicator lights illuminated, indicating that the machine started normally; only the display was malfunctioning. This suggested a graphics card issue. After powering down the machine, the graphics card was reseated. After reseating, the machine started and the monitor displayed normally, resolving the malfunction. Two days later, the hospital reported the same malfunction. A replacement graphics card of the same version was installed, and the machine operated without further issues for some time, thus completing the repair.

Summary

Compared to other large medical devices, the structure of ultrasound equipment is relatively lightweight and simple, making it relatively easy to identify the cause of malfunctions. However, quickly and accurately locating the source of a machine malfunction requires the repair engineer to have a deep understanding of the machine's structure and operating principles, along with some basic computer knowledge.

 

 

 

 

 

 