scroll down

Rencheng Repair Case Study: My Lab Twice Navigation Function Failure Repair

Release time:

2025-03-12 09:40

Machine Type: Bestsonic Color Ultrasound Doppler Diagnostic Instrument

Machine Model: MyLabTwice

1. Malfunction Phenomenon:

Navigator function failure. As shown below:

32602f2975379c9db36852dc8365f899.jpg

2. Malfunction Analysis and Cause:

Based on the malfunction phenomenon, it is first suspected that the software authorization has expired, causing this function to fail. By pressing the menu button to Enter the system information to check whether the software authorization has expired, it was found that the software authorization was normal and had not expired. However, during the software check, it was found that the system hard disk storage was full. A hard drive was then used to copy the data from the machine, freeing up system hard disk storage space. After copying the data, the machine was restarted, and the Navigator function was found to be normal, and the malfunction disappeared. Therefore, it can be determined that the cause was the full system hard disk storage space.

 

3. Malfunction Handling:

Perform software maintenance and clean up system hard disk storage space.

 

4. Repair Summary:

    During the repair process of this malfunction, it is necessary to be familiar with the system structure and system function principles of the machine. Multiple causes of malfunction can occur with one malfunction phenomenon. Rational analysis should be used to find the accurate malfunction point and thus eliminate the malfunction to ensure the normal use of the machine.

 

 

 

 

 

The pictures are from the internet. If there is any infringement, please contact the editor for deletion.

The above articles and works are all owned by Rencheng. Please indicate the source when reprinting.