Olympus GIF-H260 gastroscope image has vertical stripe fault repair

Release time:

2025-03-12 09:40

Machine Type: Olympus Endoscope

 

Machine Model: CF-Q260AI Colonoscope
1. Malfunction Phenomenon: Vertical lines in the image
37b5f9397ff4ba72956bc61a66d51ae2.jpg
2. Malfunction Detection:
(1) Observe the external tube for any damage or destruction; aging and wear of the outer skin were found.       
(2) Observe whether the illumination fiber is damaged; no obvious damage was found.
(3) Check whether the clarity of the endoscope image, field of view, etc., are normal; vertical lines were found in the displayed image.
(4) After disassembling the internal components, a faulty and burnt-out CCD component was found.
3. Malfunction Determination:
Long-term use resulted in aging and wear of the outer skin. The internal CCD component is faulty and burnt out.
4. Malfunction Handling:
Overhaul, replace the CCD component; replace the outer skin.
5. Post-Repair Effect:

9f51323309f3829c17d3dc455fbef9bd.png

dfed5efd230ccf0998a0109104d88126.png

6. Endoscope Maintenance Tips:
Parts of the endoscope that often leak water
(1) Outer skin of the bending section.
(2) Biopsy forceps channel, air and water supply pipes.
(3) Insertion tube, light guide, and objective lens at the distal end.

 

 

 