Rencheng Repair Case: Analysis of Philips IU22 Ultrasound Machine Startup Screen Display Abnormality

Release time:

2025-03-12 09:40

b1f410d5bf204d0f668ccbc0ac7074d6.jpg

Machine Type: Philips Color Doppler Ultrasound Diagnostic Instrument

Machine Model: Philips IU22

 

Fault Case 1

1. Fault Phenomenon

When the IU22 ultrasound diagnostic instrument is powered on, part of the screen display becomes smaller, the scroll bar stops loading at "29"; restarting the system stops at the "Initializing software" interface; restarting again, the fault phenomenon remains.

3821c4067b4bdeb2f290f0d8d35dc852.png

Fault: Part of the screen display becomes smaller

2. Fault Analysis

①Check the workstation display at the department, which is consistent with the ultrasound display, ruling out display faults;

②Open the back cover and find that every time the loading stops during startup, the UMB BIOS digital tube displays "AA", indicating that the host hardware loading is successful;

③Swap the two graphics cards, and the ultrasound display can normally Enter the "Initializing software" interface with the normal screen size, and then stop loading;

④To further Confirm the fault point, Reset the positions of the two graphics cards, and the fault phenomenon remains the same, thus basically suspecting a graphics card fault.

3. Troubleshooting

After replacing the spare graphics card,it can be powered on normallyUse,Troubleshooting

 

 

 

 

 

 

Some pictures are from the Internet. If there is any infringement, please contact the editor for deletion.