Olympus BF-260 electronic bronchoscope image abnormality
Release time:
2025-03-12 09:37
Machine Type:Olympus Electronic Bronchoscope
Machine Model:BF-260
1. Malfunction Phenomenon
Image abnormality

2. Fault Detection
Check for damage.The rubber in the bending section is aged and severely damaged;The angle wire is not in place.
3. Fault Judgment and Analysis
Long-term use has caused the rubber in the bending section to age and needs to be replaced; the angle wire is insufficient and needs to be adjusted.
4. Troubleshooting
Replace the bending rubber and adjust the angle.
5. Post-repair machine effect

6. Endoscope Tips
After daily use
(1) After use, the endoscope should be cleaned, disinfected, dried, and leak-tested strictly according to the instructions to extend the service life of the endoscope.
(2) After use, it should be immediately cleaned with cleaning solution, and the channel should be cleaned with a special cleaning brush before being soaked in the soaking solution. At any time, the soft part of the endoscope should not be held or cleaned together with hard and sharp accessories. When cleaning, please use clean and soft non-woven fabric to prevent foreign objects from contaminating the endoscope body.
(3) After cleaning, the endoscope should be blown dry, and the lens surface should be wiped clean in the same direction with lens paper to keep the lens clean and bright.
(4) Regularly check for leaks in the endoscope to prevent water ingress from causing imaging failures.