I. Operating Procedure

32a7be3ddacb1ba380a2df78c3c20c01.png

886a35b871f22f33522a10d4b983cefe.png

II. Troubleshooting

1. Machine cannot power on:

   1) Confirm that the battery power is sufficient;

   2) Confirm that the power is connected.

2. After powering on, there are no three probe selections, only one probe is displayed:

   1) Confirm that all three probes are tightly connected and locked;

   2) Check if the probe converter is tightly connected to the host and locked;

   3) Confirm that all three probes are connected to the converter.

3. No image is displayed during inspection:

   1) Confirm that the coupling agent has been applied to the probe;

   2) Confirm that the iTouch smart key has enabled one-key image optimization.

4. Imaging is weak during venous cannulation:

   Turn on the iNeedle puncture enhancement key.

5. Local unclear during inspection:

   1) Check if the inspection site is correctly positioned;

  2) Check if the iZoom intelligent full-screen zoom key is enabled, and whether the Auto Trace intelligent spectrum is in the zoom state.

III. Maintenance

 

1. Display screen and touch screen: Daily maintenance is recommended. After use,clean with a soft cloth or a professional cleaning kit.

2. Control panel (including keyboard and various function keys): After each use, wipe with a soft cloth dampened with water and wrung out.

3. Ultrasound probe: After each use, gently wipe with a soft cloth dampened with water. Do not use alcohol-based products to wipe.

4. Trackball component: The disinfection team cleans it in the first week of each month. Rotate the trackball retaining ring counterclockwise, gently remove the ball, use a soft cloth to clean the roller and the ball, and rotate the retaining ring clockwise after cleaning.

5. Dust cover: The disinfection team cleans it in the first week of each quarter. Take out the dust cover and clean it with a cleaning brush.

 

IV. Precautions

 

1. Operators must be certified.

2. Before powering on, check if the power socket and cables are safe, and if the instrument probe is connected.

3. After powering on, adjust the equipment conditions to the optimal state according to the patient's condition until satisfactory images are obtained.

4. During use, avoid coupling agent residue on the control panel, display screen, and touch screen to avoid affecting their functions.

5. After the inspection, wipe off the coupling agent on the probe surface and keep the probe clean; during operation and after the inspection, carefully place the probe and avoid excessive bending of the probe cable or dragging it on the ground to avoid damaging the probe.

6. After the inspection, press the freeze button to freeze the instrument.

7. Strictly follow the instrument and equipment maintenance system to maintain the equipment.

 

 

 

 

 

 

 