Rencheng Repair Case: Myray High-Frequency Probe 7L4A Image Shows Dark Channels
Release time:
2025-03-12 09:37
Machine Type:Mindray High-Frequency Probe
Machine Model: 7L4A
1. Malfunction Phenomenon
Image has dark channels
2. Fault Detection
(1) Observe the exterior for any damage. Complete and undamaged.
(2) On-machine detection. Detection shows that the transducer elements are generally weak, capacitance value 280~300pf, transducer is faulty.
4. Fault Handling
5. Post-Repair Effect

6. Probe Maintenance and Care
(1) Handle with care
Ultrasonic probes must be handled with care during use to avoid impacts that damage the internal chips. When finished using the probe, do not fold or tangle the probe connecting wires.
In daily use, when changing ultrasonic probes, always turn off the main power. If the host is not powered off and not frozen, directly plugging and unplugging the probe may cause damage to the internal circuit board and crystals due to a momentary surge of current. It will also accelerate the oxidation of the probe plug pins, directly affecting the normal use of the probe.