Machine Type:Olympus Electronic Colonoscope
Machine Model: CF-240I
1. Malfunction Phenomenon

2. Malfunction Detection
Observe whether there is any damage or destruction of the external tube. Good and undamaged; bending rubber leaks, the light beam brightness is insufficient, and the wire component is worn, causing deformation of the angle.
3. Malfunction Judgment and Analysis
Long-term use leads to bending rubber leakage, internal component wire component wear causes insufficient angle; light beam damage.Overhaul. Replace the light beam, bending rubber, and angle wire.5. Repair Effects After Installation:

6. Endoscope Maintenance Tips: The endoscope should be stored properly after use.(1) Cleaning: After use, the endoscope must be thoroughly cleaned, disinfected, dried, and maintained (leak test, alcohol wiping of the head without water, alcohol wiping of the endoscope body without water, oiling buttons, etc.).(2) Storage: Store in a dedicated endoscope cabinet (hanging, stable), and turn on the ultraviolet disinfection regularly.