Rencheng Repair Case: Olympus Electronic Colonoscope CF-240I Blurry Image

Release time:

2025-03-12 09:37

Machine Type:Olympus Electronic Colonoscope

Machine Model: CF-240I

1.   Malfunction Phenomenon
Abnormal Image

 

203e43fb510680c5d4a9774a02e2392c.png

2. Malfunction Detection

Observe whether there is any damage or destruction of the external tube. Good and undamaged; bending rubber leaks, the light beam brightness is insufficient, and the wire component is worn, causing deformation of the angle.

3. Malfunction Judgment and Analysis

Long-term use leads to bending rubber leakage, internal component wire component wear causes insufficient angle; light beam damage.
4. Malfunction Handling
Overhaul. Replace the light beam, bending rubber, and angle wire.
5. Repair Effects After Installation:

 

c87094d890a67c1097b91a29b1150a5c.png

 

6. Endoscope Maintenance Tips: The endoscope should be stored properly after use.
(1) Cleaning: After use, the endoscope must be thoroughly cleaned, disinfected, dried, and maintained (leak test, alcohol wiping of the head without water, alcohol wiping of the endoscope body without water, oiling buttons, etc.).
(2) Storage: Store in a dedicated endoscope cabinet (hanging, stable), and turn on the ultraviolet disinfection regularly.

 

 

 

 

 

 