Rencheng Repair Case: Olympus Ultrasound Endoscope UCT240-AL5 Image Blur

Release time:

2025-03-12 09:38

Machine Type:Olympus Ultrasound Endoscope

Machine Model:UCT240-AL5
1. Malfunction Phenomenon

The machine can be powered on normally without any error messages; however, the ultrasound image is blurry.

2. Fault Detection

Observed that the external tube is damaged or destroyed, the insertion tube is raised and deformed; the cable and outer sheath are both damaged; after machine inspection, the transducer is damaged.

3. Fault Determination and Analysis
Long-term use has caused the insertion tube to bulge and deform; the cable and outer sheath are damaged; the ultrasound endoscope transducer is damaged, causing blurry images. Replacement is recommended.
4. Troubleshooting
Overhaul. Replaced the transducer, cable, insertion tube, and outer sheath. After powering on the machine, the fault was eliminated, and the image is normal.
5. Machine Performance After Repair
6. Endoscope Maintenance Tips

The endoscope should be stored properly after use.

(1) Cleaning: After use, the endoscope must be thoroughly cleaned, disinfected, dried, and maintained (leak test, alcohol wipe without water on the head end, alcohol wipe without water on the body, oiling the buttons, etc.).

(2) Storage: Store in a dedicated endoscope cabinet (hanging, stable), and turn on the ultraviolet disinfection regularly.

 

 

 

 