Cleaning and disinfection of Siemens ultrasound equipment and transducers

Release time:

2025-03-12 09:39

Hospitals receive a large number of patients every day,which means there are also more germs. Hospitals have extremely strict disinfection and hygiene standards, with strict disinfection definitions and regulations for various items and equipment in the hospital environment. Cleaning, disinfection, and sterilization of medical devices are indispensable and important links. The quality of cleaning and disinfection work is closely related to the life, health, hospital infections, and medical care quality of patients.Ultrasound equipment, as the most common medical device in hospitals, also has a set of strict cleaning and disinfection procedures.Let's take a look at how Siemens ultrasound equipment is cleaned and disinfected?

 

ca5d7da6ceadfc0f26284095c0f360ad.jpg


 

Cleaning and disinfection of Siemens ultrasound equipment and probes are based on the Spaulding classification principle.

 

 

 

Daily cleaning of the ultrasound console

 

 

dac162774b979c8bd2c053b524debf74.png

 

Before cleaning the console, turn off the power and disconnect the power. After cleaning, wipe it dry with a clean, soft cotton cloth.

Housing: A soft cloth moistened with neutral soapy water can be used for cleaning (wring it out until it is not dripping).

Monitor: Use a monitor-specific cleaning agent to clean the monitor.

Filter: Remove the air filter at least once a week, remove dust with a vacuum cleaner, or wash it with clean water and let it air dry before reinstalling it.

 

Daily cleaning and disinfection of probes

 

7b2d44cfdf7592034330524693ecb4d3.jpg

Clean any residual coupling agent immediately after each examination.
It is recommended to use an enzyme-containing neutral detergent for cleaning; soapy water can be used for cleaning and rinsing (please turn off and disconnect the power before cleaning).
For probes that routinely contact the body surface, it is recommended to use disinfectant wipes containing dialkyldimethyl ammonium chloride for disinfection.
Clean the probe cables regularly, place them on hooks, and avoid dropping them on the ground to prevent contamination.

 

Daily cleaning and disinfection of special probes

7ce37f590535e735936fa14b2da2f62c.jpg

TEE transesophageal probe:

 

After use, remove the probe and protective cover.

Immediately rinse the probe with water for at least 1 minute. Wipe (using gauze soaked in 70% isopropyl alcohol or an approved cleaning agent) to remove mucus or biological contaminants.
Immerse the probe in Cidex OPA (orthophthalaldehyde) solution for at least 30 minutes. Do not exceed the last depth marking on the catheter.Parts of the probe that cannot be immersed, such as the control handle and connector, should be cleaned with 0.1% benzalkonium chloride.
Rinse with water, wipe with gauze soaked in 70% isopropyl alcohol or an approved cleaning agent, and hang to dry.

 

Intracavitary probe

 

Please use a disposable sterile sheath, then use disinfectant wipes containing dialkyldimethyl ammonium chloride for disinfection. Clean before disinfecting, and ensure one-person-one-disinfection.  

 

Intervention probe

 

Use a disinfected disposable probe cover; high-temperature disinfection should not be used during sterilization, low-temperature plasma, sodium hypochlorite, or glutaraldehyde immersion disinfection are recommended. After disinfection, rinse with sterile saline to avoid disinfectant residue.

 

Disinfection of ultrasound consoles and probes during the epidemic prevention period

31be31d9942d50c2b031ddacb2763a2c.jpg

The console, monitor, touch screen, control panel, and probe cables should be completely wrapped with protective films such as plastic film. Use disposable probe covers for the probe.

After each use, with the protective film in place, disinfect with 75% alcohol or 70% isopropyl alcohol. (Use wipes or spray disinfectant onto medical non-woven fabric, the humidity should be close to that of a wipe), be careful to prevent the disinfectant from flowing into the gaps in the control panel or the probe jack.

UV irradiation of the entire machine for 30 min/time.  

Finally, remove the protective film and probe cover, and repeatedly wipe with disinfectant wipes to ensure there are no dead corners. Then the treatment for the next patient can begin.

 

 

 

 

 