The correct way to use an esophageal volume probe

Release time:

2025-03-12 09:39

ACUSON SC2000 Z6Ms Transducer Cleaning and Maintenance Guide

 

 

15344abafb1a9252b33b66d3ba2f067c.png

 

Cardiac Intervention Artifact: Transesophageal Volume Probe Z6Ms

 

Finally bought the desired ACUSON SC2000, equipped with the high-end transesophageal volume probe Z6Ms, ultrasound can once again play a major role in the field of cardiac intervention, isn't it exciting to think about?

Don't worry, the transesophageal volume probe is a very precise device that requires our careful care on a daily basis. In order for this good comrade-in-arms to accompany us for a longer time, let's take a look at what matters we need to pay attention to.

 


I. Storage, Transportation and Accident Prevention Instructions

 

TEE Probe Box

fa6cc5e653b02096a429d67550239a84.png2b0861f91b72bc8d330a0757eb3c9d6e.png



TEE Probe Storage

 

The transesophageal probe is a device that is very sensitive to mechanical impact. The end is particularly sensitive to collisions, scratches, and vibrations. Extra care should be taken during loading, unloading, and storage.   

 

• Do not store in a humid environment

• Do not store in extreme temperatures or direct sunlight

• Do not store on the ultrasound system's probe holder

• Only store in the storage box or hang on a wall-mounted holder after disinfection

• When storing the transesophageal probe, ensure that the brake is engaged, the end is straight, and the cable and tube are not twisted; be sure to use the gray probe end protector to avoid collisions.

TEE Probe Transportation

• Place the probe in a biohazard bag, excluding the control handle

• Place in a safe and convenient container for transport

• The probe box can only be used to store and transport disinfected probes; un-disinfected probesshould notbe placed in the probe box!

8e93263f81d030088d2176007a6c99fe.png

TEE Probe Accident Prevention

 

• Check the probe for cracks or gaps  

• Perform electrical safety check

            Leakage Test   

            Conductivity Test  

• If the electrical safety test is not passed,  Do not use the probe

• Be sure to use a suitable bite block

 

21f270bd078f956732b37f6ed71d396e.png

II. Solution Conductivity and Leakage Current Test

Solution Conductivity Leakage Current Test Dale800B Tester

 

• This leakage current tester simulates an ultrasound system

• Pass the leakage current test to ensure that the leakage current is within the acceptable range

• The solution conductivity should be confirmed and then leakage current checked under the following circumstances:

   After each strict disinfection procedure

   After prolonged periods of non-use

   If the probe is damaged, detached, or broken

 

Schematic Diagram of Solution Conductivity and Leakage Current Test

 

0b12eb7d9a2e60a4231dbff6ee0d6e28.png

Please refer to the operating manual or call the relevant Siemens Healthcare representative if necessary

 

III. Correct Cleaning Instructions for the Probe

 


Probe Disinfection Steps

 

Step 1: To avoid electric shock and probe damage, be sure to disconnect the probe before cleaning or disinfecting

Step 2: Pre-cleaning before disinfection

 

• Remove the probe cover and wipe as shown

Only use approvedcleaning agents and disinfectants

• Use a pre-cleaning agent to remove mucus or bioburden before disinfection

Pre-cleaning before disinfection

bc8a2176679cf1799d3834b70e89d1c4.png

Step3: Disinfection process in the disinfection basin

•Fill a basin with disinfectant or saline solution and immerse the probe in the basin.

should notImmerse the control handle or connector in the solution, ensuring the control handle does not fall into the basin.

should notBeyond the last depth marker of the catheter.

•Use a highly effective disinfectant according to the manufacturer's instructions, immersion time.should notExceeding the time specified in the manufacturer's guidelines.

•Do not use hot steam or ethylene oxide, iodine or iodine-containing solvents, bleach, ethanol or methanol to disinfect or clean the probe.

•The end of the probe is very sensitive to pressure,Be extremely careful when cleaning the end.

•Do not bend the catheter into a curve with a diameter of less than 30.5cm.

•During disinfection, a solution conductivity leakage test can be performed.

 

Disinfection basin disinfection
2f6fcd60287ddc6b41cf836ef0776826.png

 

 

Step Four: Cleaning the probe connector housing and cable

 

  • Use a gauze pad dampened with isopropyl alcohol or an approved cleaning agent

  • Make sure to keep the probe housing and cable dry

  • Gauze/soft cloth wiping or air drying for 30 minutes, do not use hot air

 

Cleaning the probe connector housing and cable
de6038657556566b26ad6355f07391ec.png

 

IV. Types of Approved Chemical Disinfectants

 

Approved cleaning agents

  • Aniosyme DD1Quaternary ammonium compound PHMB

  • ENZOLEnzyme solution

  • VirexⅡ256Quaternary ammonium compound

     

Approved disinfectants

  • CIDEXPLUS Glutaraldehyde

  • CIDEX OPA Orthophthaldialdehyde

  • PI-Spray Ⅱ Quaternary ammonium compound

  • Sani-Cloth AF3 Diquaternary ammonium compound


V. Usage Tips

 

•Do not place the Z6Ms probe in the first probe interface

 • Avoid prolonged periods in Live status and/or 4D mode

 • Do not apply excessive force to the distal end connection during use; use the bending control to change the deflection

 

 

Occluder, must be used,

Port one, do not connect,

Live/4D, do not keep open for long,

Distal end, control rotation

 

 

 

 

 

 

 