Rencheng Repair Case: Repair of MEDISON MD3DC2-6 probe image with dark channel and other faults

Release time:

2025-03-12 09:41

Probe Type: MEDISON Probe

 

Probe Model:3DC2-6
1. Malfunction Phenomenon:
Poor echo, image with dark channels, specific malfunction phenomenon is as follows.
https://v.qq.com/x/page/d3256xgez63.html
2. Pre-repair Processing:
Disinfect, clean, and photograph the probe.
805fee16fdb2660bbdd7499fb0b925a7.png
3. Fault Detection:
The first step is to check the appearance and find that the probe cable is damaged.
The second step is electrical performance testing. Using the equipment to check the overall capacitance of the probe, it is found that the array element capacitance is inconsistent, and most of them are weak.
The third step is to disassemble the probe structure. The sound head is cracked, the internal oil sac is damaged, the four-dimensional structure is corroded, and the probe is severely damaged overall.
The fourth step is to check on the machine and view the actual image. The image is incomplete and has dark channels.

83e53ead1d95739f8b36801641507dfe.png11f12f34788442ea0ab77e07888eee2c.png

4. Fault Determination:
Severe damage to the sound head component leads to image loss and dark channels.
5. Repair Process:
Overhaul, replace the sound head, replace the oil sac, remake the cable, repair the four-dimensional drive mechanism, and test the machine after completion, the fault is eliminated.
6. Post-repair Effect:
ee2265cad87eadf31477ca3016651145.png
https://v.qq.com/x/page/t3256362oti.html

 

 

Ultrasound transducer; medical device