Rencheng Repair Case: OLYMPUS Olympus BF-P60 bronchoscope image has a large number of black spots and other faults
Release time:
2025-03-12 09:41
Endoscope Model:OLYMPUS BF-P60 Bronchoscope
Malfunction Phenomenon:
1. A large number of black spots on the image;
2. Bulging in multiple places on the insertion tube;
3. Insufficient angle


Malfunction Analysis:
1. Fiber aging and breakage in the image bundle;
2. Aging, wear, and bulging of the insertion tube
3. Wear of the wire and spiral tube
4. Aging and wear of the outer sheath
Repair Plan:Overhaul
1. Replace the image bundle
2. Replace the insertion tube
3. Replace the wire and spiral tube
4. Replace the outer sheath
Post-repair effect:



Endoscope Maintenance Tips:
The carrying case is only for transportation!
1. The environment inside the carrying case is not suitable for storing, cleaning, and disinfecting endoscopes.
2. Clean endoscopes should be hung vertically in a dry place with a moderate temperature.
3. The carrying case should also be cleaned or replaced regularly.