OLYMPUS GIF-p30 fiber optic gastroscope insertion tube malfunction: foreign object, deformation, blurry image. Repair needed.

Release time:

2025-03-12 09:41

Endoscope Model:OLYMPUS BF-P60 Bronchoscope

Malfunction Phenomenon:

1. A large number of black spots on the image;

2. Bulging in multiple places on the insertion tube;

3. Insufficient angle

 

 

Malfunction Analysis:

1. Fiber aging and breakage in the image bundle;

2. Aging, wear, and bulging of the insertion tube

3. Wear of the wire and spiral tube

4. Aging and wear of the outer sheath

Repair Plan:Overhaul

1. Replace the image bundle

2. Replace the insertion tube

3. Replace the wire and spiral tube

4. Replace the outer sheath