Machine Type:Mindray Abdominal Probe
Machine Model:C5-2s
1. Failure Phenomenon
Abnormal image with dark channels.
The lens was found to be damaged, coupling agent infiltrated, crystal array elements corroded and attenuated, and the transducer matching layer showed severe corrosion. The transducer was severely damaged.
Replace the acoustic lens and transducer.
Ultrasound probes must be handled with care during use to avoid impact damage to the internal chips. When the probe is not in use, the probe connecting wires should not be bent or tangled.(2) Probe Plugging and UnpluggingWhen changing ultrasound probes in daily applications, always turn off the main power. If the host is not powered off and not frozen, directly plugging and unplugging the probe may cause damage to the internal circuit board and crystals due to a momentary surge of current. It will also accelerate the oxidation of the probe plug pins, directly affecting the normal use of the probe.(3) Probe Cleaning and DisinfectionAvoid contact between the probe socket and disinfectant. If the probe socket is contaminated with cleaning agents and disinfectants, it will directly corrode and accelerate the oxidation of the probe socket pins. In short, maintenance is an important aspect of ultrasound probe use in medical equipment. The effective accuracy of the ultrasound probe directly affects the efficiency of ultrasound examinations. Regular maintenance should be carried out to reduce damage caused by human factors and extend the service life of the ultrasound probe.