Rencheng Repair Case: FUJINON Endoscope Image Screen Flickering

Release time:

2025-03-12 09:38

Machine Type: FUJINON Endoscope

 

Machine Model: EG-250WR5

1. Malfunction

 Blurry image with snow, blue screen.

20b04ae2c24bb9e9f3fc81896acb3d3d.png

2. Malfunction Detection

Observe the external tube for damage, wrinkles on the light guide sheath, and bending of the forceps tube.

3. Malfunction Judgment and Analysis

Long-term use has caused bending of the forceps tube, wrinkles in the light guide sheath, and aging of the internal angle wire of the insertion tube, resulting in improper angle. Angle wire detachment, rust; residual tissue protein on the surface of the insertion tube is dehydrated and adhered to the outer sheath during disinfection, accumulating over time to form crystals, causing damage to the CCD component, and water leakage in the large knob.

4. Troubleshooting

Overhaul; Replace CCD components, forceps tube, light guide, large knob, rubber; adjust the angle. After replacement, the malfunction was eliminated.

5. Post-Repair Effect

 

958d82d041fe0830330ef3bb263a048f.png

 

6. Endoscope Maintenance Tips

Endoscopes should be properly stored after use.

(1) Cleaning: After use, endoscopes should be thoroughly cleaned, disinfected, dried, and maintained (leak testing, wiping the distal end with anhydrous alcohol, wiping the body with anhydrous alcohol, oiling the buttons, etc.).

(2) Storage: Store in a dedicated endoscope cabinet (hanging, stable), and turn on the ultraviolet disinfection regularly.

 

 

 

 

 

 

 

 

 

 

 

This information belongs to Rencheng, please indicate the source when reprinting.