Rencheng Repair Case: Dark streaks in the image of Mindray 3C5A ultrasound
Release time:
2025-03-12 09:38
Machine Type:Mindray Abdominal Probe
Machine Model:3C5A
1. Malfunction Phenomenon:Image has dark channels.

2. Fault Detection:
(1) Observe the external damage, and find that the probe cable and the casing are both damaged.
(2) Instrumental detection revealed that the lens is bulging, and the crystal array elements are corroded by the coupling agent, resulting in performance attenuation. The probe is damaged and replacement is recommended.
3. Fault Judgment and Analysis:
Long-term use has resulted in damage to the cable and casing. Inspection revealed probe damage, and the image shows dark channels.
Fault Handling: Intermediate repair, replace the probe, casing, and repair the cable. The image is normal after powering on.
4. Post-Repair Effect:

5. Probe Maintenance Tips
(1) Ultrasonic probes are expensive components. During use, they must be handled with care, and the probe head must not be impacted. During transportation and storage, use the original manufacturer's probe packaging box or wrap the probe head in a soft cloth.
(2) When installing and removing the probe, turn off the main power of the machine first, then carefully operate.
(3) Avoid contact with organic solvents, otherwise the probe may crack and be damaged due to contact with organic solvents. However, alcohol-based organic solvents can be used. Remember not to use acetone for wiping.
(4) Protect the lens surface. Once the lens surface is damaged, the coupling agent easily enters the probe, damaging the piezoelectric crystal. Therefore, sharp objects are not allowed to pierce the lens surface during use.
(5) Use a non-corrosive coupling agent. The ultrasonic coupling agent used has no irritating effect on the skin and is non-oily, because many probes' acoustic lenses are easily corroded and damaged by oily substances.
This information belongs to Rencheng Medical. Please indicate the source when reprinting.