Pentax Electronic Gastroscope EG29-i10 Electrical Department Water Ingress Case Analysis

Release time:

2025-03-12 09:39

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.

 

 

 

 

 

44640b97d616e9de67b234545befa8f4.jpg

During EG29-i10 endoscope use, the color is abnormal and purple.

 

The metal casing was opened, and obvious water stains were found on the driver board. The driver board plug was corroded and rusty. After cleaning the driver board solder joints and plugs, and re-soldering the wires, the image returned to normal.

fdf83a94408285a563fd043b66f386bf.jpg

In this case, water ingress into the electrical department was due to the loosening of the electrical department sheath, causing water immersion during cleaning and disinfection!

Image problems were reported promptly, preventing serious consequences such as CCD short circuit and burning.

Points to note: During endoscope use, pay attention to leak testing before and after use. If leakage is detected, stop using it immediately and report for repair. Water ingress into the electrical department can cause image flickering or black screen, and short circuit damage to the driver board CCD! Water ingress into the operating department can cause stiff angles, broken steel wire reels, no angle, and water ingress into the bending department can cause CCD short circuit and stiff steel wire!

 

 

 

 

 

This information originates from the internet. Please contact us for removal if there is any infringement.