Analysis of Ultrasound Equipment Failure Repair Cases
Release time:
2025-03-12 09:39
1、Monitor Black Screen
For a fixed black screen, it generally appears as a black screen after powering on. First, we observe the status of the instrument. Most of the current ultrasound operation panels and main processor modules will have some indicator lights or backlights. Confirm whether some light-emitting diodes are lit. If there is no response, check the main power supply directly. If the indicator light on the processor or operation panel is lit, we can divide it into two steps: First, check whether the voltage of the power supply to the main processor and the display is normal (there are usually some test points on the power supply that can be used for judgment). If it is normal, rule out the power supply, and the remaining considerations are the display itself and the processor. Second, for the display, you can consider using an external display (the voltage range is generally 110~240V), directly connect to the power supply, pay attention to the signal cable (commonly used VGA and DVA cables), and for the processor, generally consider the graphics card and memory.
For intermittent failures, the external method can be used to reversely troubleshoot. Taking the Philips HDI-4000 ultrasound as an example, after the machine is working normally, turn it off and then turn it on. The host panel indicator light is on, the keyboard control is effective, but the display is black. Analysis and inspection: The host keyboard control is effective, indicating that the host is working normally, the display is black, and the possible reasons are: ① The host does not output a video signal to the main display, causing the display to be black. ② Display failure: causes the screen to become black. ③ The display requires a +12V DC working power supply, which is provided by the host. If the host cannot supply a +12V DC working power supply, the display will also become black. Starting with the simplest check, first measure the power supply terminal of the display, and the voltmeter shows that the voltage is 0V. This means that the black screen of the display is caused by the lack of DC working voltage, so check the power supply board PsM, and find that the +12V switching tube that supplies the working voltage of the display in the host switching power supply box is open-circuit and damaged, resulting in no normal voltage output. After buying a transistor of the same model in the market and replacing it, there is a +12V output, the power supply is normal, the display black screen fault is eliminated, and the ultrasound works normally.


2、The instrument can be turned on normally. But abnormal during use
3、Instrument starts up with a blue screen

4、The instrument has no response when turned on
Taking Philips ENVISOR as an example, the fault is that there is no response when the machine is turned on. Observation shows that the indicator light on the keyboard is not lit, the main processor light is not lit, and the internal fan does not rotate. Check that there is no voltage at the four AC output ports of the main power supply, and a row of DC output test points and light-emitting diodes are not lit, and the main power supply AC indicator light is off. However, after re-plugging the main power cord, the main power supply AC indicator light is on, and the AC power supply to the main processor is normal. At this time, pressing the power button still has no response, but at the same time, the main power supply AC indicator light is off. Consider that there may be a short circuit phenomenon inside the processor, causing the main power supply to cut off the power supply for self-protection. After inspection, it was found that the small power supply inside the processor was faulty, and it was normal after replacement.
This information comes from the Internet. Please contact us to delete it if it infringes on your rights.
Previous Page