scroll down

Rencheng Repair Case: Olympus LF-TP bronchoscope severe air leak and other faults repair

Release time:

2025-03-12 09:41

Machine Type: Olympus

 

Machine Model: LF-TP Bronchoscope
1. Malfunction Phenomenon:

 

Severe air leakage in the bending section

 

2. Malfunction Analysis and Causes:

 

Two pinholes in the outer sheath caused severe air leakage in the bending part of the endoscope

 

 

 

3. Malfunction Handling:
Replace the endoscope outer sheath, perform water and air leakage test, and the malfunction disappears

 

4. Post-Repair Effect:

 

 

 

5. Endoscope Maintenance Tips

 

Endoscope Cleaning and Disinfection Precautions

(1) Do not place the endoscope and treatment accessories overlapping in water

(2) Do not disassemble the side leak device interface, waterproof cap, and cleaning and disinfection device in water

(3) Be sure to remove the EOG cap during cleaning and disinfection

 

This article is an original work of Rencheng Medical

 

Medical devices; fault repair;